Combining MALDI-2 and transmission geometry laser optics to achieve high sensitivity for ultra-high spatial resolution surface analysis


AUTHORS

Spivey EC , McMillen JC , Ryan DJ , Spraggins JM , Caprioli RM , . Journal of mass spectrometry : JMS. 2019 4 1; 54(4). 366-370

ABSTRACT

A transmission geometry optical configuration allows for smaller laser spot size to facilitate high-resolution matrix-assisted laser/desorption ionization (MALDI) mass spectrometry. This increase in spatial resolution (ie, smaller laser spot size) is often associated with a decrease in analyte signal. MALDI-2 is a post-ionization technique, which irradiates ions and neutrals generated in the initial MALDI plume with a second orthogonal laser pulse, and has been shown to improve sensitivity. Herein, we have modified a commercial Orbitrap mass spectrometer to incorporate a transmission geometry MALDI source with MALDI-2 capabilities to improve sensitivity at higher spatial resolutions.